ANALOG BENCH ATE
APPLICOS DRS20 Module
A dual-channel reference source offering ultra high precision and stability
The DRS20 provides a +/-10 V output. The output current can be up to 10 mA and the 4 wire connection provides a 1 µV/mA load regulation. The high precision and stability is achieved by using a temperature controlled voltage reference. All internal nodes are calibrated using a built in 24-bit precision DVM that continuously checks the internal nodes relative to this reference. Therefore the accuracy is only determined by the temperature controlled voltage reference. This ensures a negligible drift when sourcing over a long period of time. The same DVM can be used to measure external voltages with extremely high precision.
The output has a very low noise level. Special attention has been paid to low frequent noise (<10 Hz) since this is very difficult to filter away afterwards. The unit is very suitable as the reference source for data converter testing.
General | |
---|---|
Channels | 2 |
Resolution | 20 bit |
Output range | -10 V to +10 V |
full accuracy output current | 10 mA |
Maximum Output current | 30 mA typical |
Output configuration | 2 or 4-wire |
Settling time for full accuracy | 20 ms (*) |
(*) The settling time is programmable to allow a trade-off between settling time and accuracy.
Accuracy | |
---|---|
Accuracy | ±(25 µV + 10 ppm of Value) |
Linearity (typical) | 2 ppm |
Temperature drift (typical) | ±(2 µV + 1 ppm of value)/ºC |
Load regulation (typical) | 1 µV/mA |
Output noise (typical) | 5 µVrms (DC to 100 kHz) |
0.1 Hz to 10 Hz noise (typical) | 4 µVpp |
Voltage measure resolution | 2.7 µV |
Calibration output | |
---|---|
Connector style | SMB |
Output voltage | 7.2 V |